Please use this identifier to cite or link to this item: https://knowledgecommons.lakeheadu.ca/handle/2453/536
Title: An integrated ultrasound transducer driver for HIFU applications
Authors: Wong, Wai
Keywords: Ultrasound transducer;Amplifier topologies;High intensity focused ultrasound (HIFU)
Issue Date: 22-Jan-2014
Abstract: This thesis proposes an MRI-compatible integrated CMOS ampli er that is capable of di- rectly driving an ultrasound transducer for HIFU applications. The output stage of the integrated amplifier operates in class DE mode with its output directly connected to a shunt capacitor and an ultrasound transducer without the need for an inductor. This design was simulated with Spectre simulator using the 0:8 m 5/20 V CMOS process data available from Teledyne-DALSA Semiconductor. The proposed integrated ampli er has an e ciency of 80% with 1 W of output power at 1 MHz and achieves an acceptable level of third harmonic. A layout of the integrated ampli er was prepared. The integrated ampli er occupies a die area of approximately 2:5 mm by 1:6 mm including input-output pads.
URI: http://knowledgecommons.lakeheadu.ca/handle/2453/536
metadata.etd.degree.discipline: Engineering : Electrical & Computer
metadata.etd.degree.name: M.Sc.
metadata.etd.degree.level: Master
metadata.dc.contributor.advisor: Christoffersen, Carlos
Appears in Collections:Electronic Theses and Dissertations from 2009